1. Scheme of the patient-specific computational workflow to predict the risk of device-related thrombus (DRT) formation after left atrial appendage occluder (LAAO) implantation. A: computerized tomography (CT) scan acquisition of whole left atrium (LA) and ultrasound (US) study with Doppler measurements at mitral valve (MV) level. B: 3D LA segmentation and model generation where finite volume analysis was performed. C: blood flow velocities and in silico indices like endothelial cell activation potential (ECAP) estimated from personalized computational fluid dynamics simulations. D: the risk factors predicting the presence of DRT were low velocities (< 0.20 m/s), and stagnated flow next to the device surface as well as high ECAP values (indicative of complex blood flow patterns and low wall shear stress).